The CardiArc SPECT System

General Specifications

Room size
–Minimum 6 x 7 ft. (1.8 x 2.1 m)
–Typical 8 x 10 ft
Camera Footprint 46.3 x 54.8 in. / 117 x 139 cm
Total Camera Weight 1,460 lbs. / 664 kg
Floor Loading (without patient) 39.6 lbs./in. sq. / 2.8 kg/cm sq.

Patient Imaging Chair

Maximum Patient Weight 400lbs. / 182 kg
Vertical Travel 7.1 in. / 18 cm

Power Requirements

120/220V @ 15A; dedicated circuit breaker
Power consumption  
–Maximum 350 watts
–Typical less than 100 watts

Environmental

Operating Temperature
–Range 60ºf to 80ºf / 16ºc to 27ºc
–Maximum variation 10ºf / 5.5ºc per hour
Humidity 30% - 80% non-condensing
Max heat dissipation 2,400 BTU / hr. (853 from computer)